The Additel 762 Automated Pressure Calibrator is unlike any other pressure calibrator on the market. This revolutionary product is a complete turnkey solution for automation of pressure calibration work up to 10,000 PSI.
Designed for use in both the field and the laboratory, the portability and accuracy of this state-of-the-art product will quickly become the favorite go-to calibrator for lab personnel and field technicians alike.
With fully automated support for calibration of pressure transmitters, switches, dial and digital gauges and sensors, including HART/PROFIBUS devices in conjunction with a fully integrated task feature, data collection and Wi-Fi connectivity, we had our customer’s needs in mind when designing our most capable pressure calibrator to date.
Pressure Range |
15-10,000 psi (1-700 Bar) |
Range Selection |
Manual 3K psi, Manual 10K psi or Auto-range |
Maximum External Load Capacity |
Max: 80 ml @ 700 Bar, 50 ml reccomended |
Reservoir |
Max: 350 ml, push-on reservoir cap, built-in filter |
User Interface |
Color touch screen and keypad operation |
Display |
7” TFT touch screen 800 x 480 color |
Enclosure IP Rating |
IP31 |
Power |
Dedicated lithium battery or power adapter |
Battery |
Rechargeable Li-Ion battery, typically 12 hours of operation, less than 5 hours recharge. |
Weight |
28 lbs (12.8 kg) without media |
Media |
Sebacate oil |
Size |
11.81 x 8.66 x 7.56 in (300 x 220 x 192 mm) |
Communications |
USB, LAN, Bluetooth, Wi-Fi and Ethernet |
HART Communication |
Read, configure and calibrate HART devices - DD files updated periodically |
Data Storage |
>8 GB |
Data Logging |
Up to 1,000,000 readings (data and time stamped) |
Task Documentation |
Up to 1,000 tasks can be stored with data |
Automation Functions |
Switch test, auto step, leak test |
User Interface Localization |
English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian |
Pump Life |
> 1,000,000 cycles |
Environmental Specifications |
Operating temperature: 32 °F to 122 °F( 0 °C to 50 °C) Storage temperature: -20 °C to 60 °C ( -4 °F to 120 °F) Humidity: <90%, non-condensing |
Certification |
ISO 17025 accredited certificate of calibration with NIST-traceable data |
Compliance |
CE |
Software Compatibility |
CE Software Compatibility ACal, Additel Land and Additel Link for access via mobile application |
Warranty |
1 year |
Hose & Filter End of Life |
The estimated End of Life (EOL) expectancy for all hoses and filters (pneumatic and hydraulic) is approximately 10 years and should be replaced at the first sign of wear or damage. |