Test voltage |
100V, 250V, 500V, 1000V, Max.1000V |
Measuring ranges |
2.000/20.00/200.0M? (Auto-ranging), 20.00/200.0/1000M? (Auto-ranging), 20.00/200.0/2000M? (Auto-ranging), 0-1000V (goes up by 1V) |
Accuracy |
±2%rdg±6dgt (2.000/20.00M?) ±5%rdg±6dgt (200.0M?) ±2%rdg±6dgt (20.00/200.0M?) ±5%rdg±6dgt (1000M?), ±2%rdg±6dgt (20.00/200.0M?), ±5%rdg±6dgt (2000M?), ±5%rdg±5dgt |
Rated current |
1.0-1.2mA @0.1M? 1.0-1.2mA @0.25M?, 1.0-1.2mA @0.5M?, 1.0-1.2mA @1M? |
Short circuit current |
1.5mA max |
Loop impedance |
|
Function |
|
LOOP ATT |
L-PE/L-N (3wire), L-PE(2wire) |
LOOP HIGH |
L-PE (0.01?Res), L-PE (0.001?Res) High test current(25A), L-N/L-L |
Rated voltage |
100-260V (50/60Hz), 48-260V (50/60Hz), 48-260V (50/60Hz), 100-260V (50/60Hz), 48-500V (50/60Hz) |
Impedance range |
20.00/200.0/2000? (Auto-ranging), 20.00/200.0/ 2000? (Auto-ranging), 2.000?, 20.00? |
Accuracy |
±3%rdg±6dgt , ±3%rdg±10dgt, ±3%rdg±4dgt, ±3%rdg±25m?, ±3%rdg±4dgt |
Nominal test current at 0? external loop: Magnitude/ Duration at 230V |
L-N:6A/60ms N-PE:10mA, L-PE:15mA, 20?: 6A/20ms 200?: 0.5A/20ms 2000?: 15mA/500ms, 25A/20ms, 6A/20ms |
PSC/PFC |
|
Range |
2000A/20kA (L-N(PSC)/ L-PE(PFC)), 2000A/20kA (PFC), 2000A/20kA (PFC), 2000A/50kA (PFC), 2000A/20kA (PSC) |
Accuracy |
PSC/PFC accuracy is derived from measured loop impedance specification and measured voltage specification |
RCD |
|
Rated voltage |
100-260V(50/60Hz) |
Function |
x1/2, x1,x5,Ramp,Auto,Uc , 6/10/30/100/300/500/1000mA/variable |
RCD type |
AC(G/S), A(G/S), F(G/S), B(G/S), EV |
Trip current setting |
|
x1/2, x1,Uc |
10/30/100/300/ 500/1000mA(G) 10/30/100, 10/30/100/ 300/500mA/ 300/500(S), 10/30/100/ 300/500mA, 10/30/100/ 300mA, 6mA(x1 only) |
x5 |
10/30/100mA 10/30/100mA, 10/30/100mA, 10/30mA, |
Ramp |
10/30/100/ 300/500mA 10/30/100/ 300/500mA, 10/30/100/ 300/500mA, 10/30/100/ 300mA, 6mA |
Accuracy |
|
Trip current |
|
x1/2 |
-8% - -2%, -10% - 0%, -10% - 0%, -10% - 0% |
x1 |
+2% - +8%, 0% - +10%, 0% - +10%, 0% - +10%, 0% - +10% |
x5 |
+2% - +8%, 0% - +10%, 0% - +10%, 0% - +10% |
Ramp |
-4% - +4% -10% - +10%, -10% - +10%, -10% - +10%, -10% - +10% |
Trip time |
|
x1/2 |
2000ms(G/S):±1%rdg±2ms |
x1 |
550ms(G):±1%rdg±2ms,1000ms(S):±1%rdg±2ms, 10.5s:±1%±2? |
x5 |
410ms(G/S):±1%rdg±2ms |
Continuity |
|
Range |
20.00/200.0/2000? (Auto-ranging) |
Open circuit voltage(DC) |
7-14V |
Measuring current |
|
200mA |
>200mA or more(2? or less) |
15mA |
15mA±3mA(short-circuit) |
Accuracy |
±2%rdg±8dgt |
Volts |
|
Range |
300.0/600V(Auto-ranging) |
Measuring ranges |
|
Volts |
2-600V |
Frequency |
45-65Hz |
Accuracy |
|
Volts |
±2%rdg±4dgt |
Frequency |
±0.5%rdg±2dgt |
Phase Rotation |
|
Rated voltage |
48-600V(50/60Hz) |
Remarks |
Remarks Correct phase sequence: are displayed “1.2.3” and mark Reversed phase sequence: are displayed “3.2.1” and mark |
Earth |
|
Range |
20.00/200.0/2000?(Auto-ranging) |
Accuracy |
±2%rdg±0.08?(20.00?) ±2%rdg±3dgt(200.0/2000?) |
Applicable standards |
IEC 61010-1 CAT IV 300V ,CAT III 600 V Pollution degree 2 IEC 61010-2-034 IEC 61557-1,2,3,4,5,6,7,10 IEC 60529(IP40) IEC 61326(EMC) |
Power source |
LR6 (AA) (1.5V) ×8 |
Communication interface |
USB, Bluetooth®5.0,Android™5.0 or more, iOS 10.0 or more ?1 |
Dimensions / Weight |
136(L) × 235(W) × 114(D) mm / Approx. 1300g (including batteries) |
Included Accessories |
Main test lead?2 7281(Test leads with remote control switch) 7246(Distribution board test lead) 7228A(Earth resistance test leads) 8041(Auxiliary earth spikes[2 spikes/1 set]) 9084(Soft case) 9142(Carrying case) 9151(Shoulder strap) 9199(Shoulder pad) LR6(AA) × 8, Instruction manual, Calibration certificate |
Optional Accessories |
8212-USB(USB adaptor with "KEW Report(Software)'') 8259(Adapter for measurement terminal) 7272(Precision measurement Cord set) 8017A(Extension prod long) |