Power Analyser up to 8ch
Model No: PW8001

Catalogue

The PW8001 sets a new benchmark for complete power analysis solutions with Hioki being the only manufacturer that offers industry-leading power analyzers and a wide portfolio of high-accuracy current sensors both developed and manufactured by Hioki. This enables to align their performance and results in more accurate and stable measurements from DC to high frequencies and low power factor measurements thanks to functions like the automatic phase shift correction.

Measurement lines

1-phase-2-wire, 1-phase-3-wire, 3-phase-3-wire, 3-phase-4-wire

No. of PW8001 input units

Max. 8 units (mix and match)

Type of input unit

U7001 2.5 MS/s INPUT UNIT U7005 15 MS/s INPUT UNIT

Measurement frequency band

U7001: DC, 0.1 Hz to 1 MHz U7005: DC, 0.1 Hz to 5 MHz

Sampling

U7001: 2.5 MHz, 16-bit U7005: 15 MHz, 18-bit

Data update rate

10 ms, 50 ms, 200 ms

Accuracy for power (U7001)

50/60 Hz: 0.02% of reading + 0.05% of range DC: 0.02% of reading + 0.05% of range 50 kHz : 0.4% of reading + 0.1% of range

Accuracy for power (U7005)

50/60 Hz : 0.01% of reading + 0.02% of range DC : 0.02% + 0.03% of reading of range 50 kHz: 0.15% of reading + 0.05% of range

Measurement range

Voltage: 6 V, 15 V, 30 V, 60 V, 150 V, 300 V, 600 V, 1500 V Current: (probe1) 100 mA to 2 kA, (probe2) 100 mA to 50 kA (Range configuration changes depending on the current sensor used, U7001 only for Probe2.)

Measurement parameters

voltage (U), current (I), active power (P), apparent power (S), reactive power(Q), power factor (λ), phase angle (φ), voltage frequency(fU), current frequency(fI), efficiency (η), loss (Loss), voltage ripple factor (Urf), current ripple factor (Irf), current integration (Ih), power integration (WP), voltage peak (Upk), current peak (Ipk) - Harmonics measurement: maximum analysis order 500th - Waveform recording: recording capacity 5M words × ( [voltage/current] ) × - No. of channels + motor wav

Calculation function

Efficiency and loss calculations, User-defined calculations*, Delta conversion, Current sensor automatic phase shift calculation

External interface

USB flash drive, LAN, GP-IB, RS-232C, external control, optical link*, BNC sync.*, CAN or CAN FD*

Power supply

100 V to 240 V AC, 50/60 Hz, 230 VA

Dimensions and mass

Approx. 430 mm (16.93 in) W × 221 mm (8.70 in) H × 361mm (14.21 in) D, Approx. 14kg (493.84 oz)

Included accessories

Power cord ×1 - Instruction Manual ×1 - GENNECT One (PC applications) CD ×1 - D-sub 25-pin connector ×1 (PW8001-02, -05, -12, -15 only)
No datas found
related products